COM-3400 Specification
Measurement system | The Gerdien capacitor method |
Measurement | There are two kinds of measurement modes, automatic measurement and manual
measurement, and it chooses according to a use and can measure. ■ Automatic measurement : A positive ion and negative ion are measured automatically and the display output of the measured value is carried out. ■ Manual measurement : A positive ion and an anion are measured individually and the display output of the measured value is carried out. |
Time base range | Positive ion, Negative 0-5,000,000 1 piece/cc Measurement range : R1= 0 - 8,000 ion/cc R2= 0 - 50,000 ion/cc R3= 0- 500,000 ion/cc R4= 0 - 5,000,000 ion/cc |
Display for indication | Character 16-figure liquid crystal of two lines The display for indication of a STN system liquid crystal, and back light with an ON/OFF control facility |
Clock function | The built-in clock which had a function of - part and a second at the time of year, moon, and day - is used. |
Memory function | In order to carry out a memory, there are two kinds, automatic mode and a manual mode. 500 data can be saved. ■ Automatic memory : A memory is carried out with arbitrary time preset values. A preset value is 1, 2, 3, 4, 5, 10, 15, 20, 30, and 60 minutes. (It synchronizes with a built-in clock 5 minutes or more) ■ Manual memory : A memory is carried out to the arbitrary time under measurement by "MEMORY switch" setup. |
Analog output | ■ Analog output 1 (OUT1) : The measurement range under measurement, (0-1V) An analog recorder, a data recorder ■ Analog output 2 (OUT2) : The output range of R1, R2, R3, and R4 is set up. (0-1V) An analog recorder, a data recorder |
Serial communication (RS-232C) | Data is outputted to external apparatus, such as a personal computer, in
real time. It can be used for an external monitor etc. A set period is 1, 5, 10, 15, 20, 30, and 60 seconds. |
Personal computer output | The data which carried out the memory can be outputted to a personal computer in the Excel form. |
Printer output | Drawing of data and digital printing are performed to a thermal printer. |
Electric mobility | 5.00 - 0.16 cu/V・sec (125 Steps Electric mobility) |
Power supply | DC12V ±10% (Power supply protection fuse 2A) |
Temperature sensor | Time base range: 0 - 50 ℃ Accuracy: 1 ℃ |
Humidity Sensor | Time base range: 5% - 95% Accuracy: 5% |
Ion measurement | 5 ~ 35 ℃ 85%RH ≧ (Don't dew) |
Main part | size: 130mm x 260mm x 340mm (A projection thing is not included) weight : About 5 k |
Detector | size:φ70mm x 435mm (A projection thing is not included) weight : About 4 kg |
Apparatus composition | Main part, Detector, Connecting cable (5 m) = 2, Cable for temperature humidity (5 m), Data collection display PC software, RS-232C connecting cable, Main part aluminum suitcase, Detector aluminum suitcase, AC power supply Adapter(use power-supply range: AC100-240V), Operation manual |
Option | ■ Make measurement box |